Home Jinjiang Electronics Receives NMPA Approval for Second-Generation Pulsed Field Ablation Catheter

Jinjiang Electronics Receives NMPA Approval for Second-Generation Pulsed Field Ablation Catheter

Apr 22, 2025 12:13 CST Updated 12:13
Jinjiang Electronic

Developer of electrophysiology products


Image

Award Selection Postponed! Cardiovascular Enterprises Act Now! First Global Cardiovascular Conference!



Heart Future

On April 21, 2025, the National Medical Products Administration approvedSichuan Jinjiang Electronic Medical Device Science and Technology Co., Ltd.(Abbreviation: "Jinjiang Electronic")" of "PulsedFA®FocalPointSingle-Use Magnetic Localization Pressure Monitoring Pulsed Field Ablation Catheter"(China Medical Device Registration No. 20253010788) Innovative Product Registration Application."

Image

As of now, the cumulative number of innovative medical devices that have been listed in China346 items


This product consists of a tip, tube body, handle, and connecting cable. It is used in conjunction with the cardiac pulse ablation instrument produced by our company and is mainly used forTreatment of Paroxysmal Supraventricular Tachycardia

Single-use magnetic positioning pressure monitoring pulsed electric field ablation catheter throughAblation Tip StructureThe innovative design enables the precise application of high-voltage pulse energy sources on target tissues. The product adopts a multi-electrode configuration in localized areas, enhancing the adaptability of the ablation tip to various complex and delicate tissue structures.Can display catheter position and tip pressure monitoring values in real time., effectively improving surgical efficiency and reducing the risk of cardiac perforation complications in patients.

This is Jinjiang ElectronicApprovedUpCity'sThe Second Self-developedPFAPulsed Field Ablation Catheter(The first paragraph isPulsedFA® Single-use Cardiac Pulsed Field Ablation Catheter), and both are selectedNMPACreation of the Green ChannelNew Product: The launch of this catheter marks an important expansion of Jinjiang Electronic's product pipeline in the field of tachyarrhythmia.

Image
▲ Image source: Jinjiang Electronic Official WeChat
So far, Jinjiang Electronic has launched 3 products.PFAInnovative products, and theirLEAD-Mapping®3D Cardiac Electrophysiology Mapping SystemHighly integrated, building a moreFor completionThe 3D Impulse Ablation Platform of Jinjiang Electronic.

# Research and Development Background
The population suffering from atrial fibrillation is vast and is significantly affected by the accelerating aging of the population.It is estimated that by 2025, the number of people with atrial fibrillation will reach 22.666 million., with a compound annual growth rate of 2.85% from 2021 to 2025, indicating a high demand for surgical procedures.
In the field of cardiac ablation, including "Ice”、“Fire”、“Electricity"Cardiac Electrophysiology Solutions with Three Types of Ablation Energy – 'Fire Element'"Radiofrequency Ablation"Ice Element"CryoablationAnd "Electric Elements"Pulsed Field Ablation (PFA)
In the past decade, traditional radiofrequency ablation and cryoablation as interventional treatments for atrial fibrillation have become first-line treatment options, butTraditional ablation methods still have issues with pulmonary vein stenosis and the left atrium.Esophageal Fistula, severe surgical complications such as phrenic nerve injury and cardiac tamponade. Therefore, how to carry out more safely and effectivelyAtrial Fibrillation Catheter Ablation SurgeryIt is still one of the pain points困扰 clinical practice.
Different from the other two traditional ablation methods,PFA ablates myocardial tissue in a non-thermal manner with tissue specificity, ablating myocardial tissue while preserving surrounding tissue structures, avoiding conductive thermal injury and barotrauma.
According to Frost & Sullivan's forecast,China's PFA Market Size to Reach 1.3 Billion Yuan by 2025, and maintain rapid growth,Expected to reach 16.3 billion yuan by 2032, with a compound annual growth rate of 43.73% during the period. The proportion of China's PFA market size in the overall electrophysiology device market size is also expected to continuously increase, rising from 8.18% in 2025 to 38.87% in 2032.

图片


# PulsedFA®FocalPointCatheter

PulsedFA® FocalPoint It is a product used for cardiac pulsed electric field ablation (PFA) innovative medical device, mainly used for treating cardiac arrhythmias and other heart diseases.

Image

Product Features

  • Multifunctional Integration:PulsedFA® FocalPoint CatheterIntegrated 3D modeling, mapping, and pulsing/Integrated with radiofrequency ablation and pressure monitoringThis multifunctional design reduces the number of catheters used and punctures during surgery, improving the convenience and safety of the operation.

  • Magnetic Positioning Navigation:The catheter, based on magnetic navigation technology, is able toReal-time display of catheter position and tip pressure monitoring values. This technology not only improves the precision of the surgery but also helps the surgeon more accurately select the ablation target.

  • Tissue Selectivity and Safety:The ProductUtilizing the tissue selectivity of pulsed electric fields, it is able toPrecisely perform ablation on myocardial cells, without damaging the surrounding tissue. Moreover, its ablation process is non-thermal ablation, avoiding heat-related injuries.

  • Ease of Operation:The catheter is made ofCircular Tip Design, which conforms to the operating habits of most doctors and has virtually no learning curve. The surgical procedure is simple, allowing doctors to quickly grasp the key points of operation.

Clinical Application and Advantages

  • High Efficiency:The catheter can simultaneously perform modeling, mapping, ablation, and validation, reducing operation time and complexity.

  • Low Radiation: During the surgical procedure, this catheter requires only a minimal amount of radiation during the interatrial septum puncture.XLine exposure, and close to zero radiation exposure for the rest of the time, significantly reducing radiation exposure for both medical staff and patients.

  • Patients experience good: Due to its non-thermal ablation and tissue selectivity, patients experience...Almost no muscle twitch reaction, with good clinical experience.


Relevant Clinical Trials

Electrophysiology Team of West China Hospital, Sichuan UniversityThe first human exploratory study was launched, and2023YearPublish the main research results inTop international journal in the field of arrhythmia, "JACC:EP, the study isPFAProvides a new perspective in the clinical application of interventional treatment for arrhythmia.

Image

This study led by Huaxi Hospital is a single-center, single-arm, prospective clinical trial, aiming toEvaluation PulsedFA® FocalPoint Catheter in TreatmentAtrioventricular Nodal Reentrant Tachycardia (AVNRT) PatientSafety and Feasibility in China.

Research Methods

  • Research Object: Included30Confirmed by electrophysiological examination asAVNRT The patients, with an average age of47.9Years.

  • Treatment method: All patients were treated with PulsedFA® FocalPoint Catheter for pulsed electric field ablation therapy.

  • Evaluation Metrics:

    Primary Endpoint:Acute Ablation Success Rate

    Secondary endpoints:Postoperative6Success Rate and Safety Events of Ablation Over the Months

Research Results

  • Acute Ablation Success Rate: All30All patients achieved acute ablation success.

  • Operation Time: The average operation time is109.1Minute,XFluoroscopy time is4.1Minute.

  • Number of Ablations: Median =8Sub-pulse Electric Field Ablation Application.

  • Follow-up Results:Postoperative6months, all patients maintained sinus rhythm and no adverse events occurred.

  • Safety:During and after the ablation process6No serious complications were reported during the follow-up period of months.

Research Conclusion

PulsedFA® FocalPoint Catheter in TreatmentAVNRT Showed good feasibility and safety in patients.All patients achieved acute ablation success and postoperative6Months of maintaining sinus rhythm, with no adverse events reported


# LEAD-Mapping®3D Cardiac Electrophysiology Mapping System

This product can be used in conjunction with the magnetic localization mapping catheter and body surface reference electrode produced by our company, which have been approved for use with this product.Emitting electrical stimulation signals and recording electrocardiogram signals during cardiac electrophysiology procedures,For cardiac electrophysiological mapping, 3D modeling, and displaying catheter positions. It can also measure intracavitary and vascular pressures. Used in medical institutions.

Image

Product Features

1. Adopting magnetic and electrical fusion positioning, with precise positioning accuracy.±1.0mm. Combined with Jinjiang Multi-Magnetic Catheter, full magnetic positioning of the catheter can be achieved.Accurately display catheter morphology;

2. With two respiratory learning technologies, respiratory compensation and gating, it can effectively shield the catheter motion interference caused by respiration, displaying the catheter motion posture in real time, stably and accurately, ensuring safe and efficient catheter operation while reducing radiation exposure.

3. The signal is true and clear,Accurately Present Characteristic Potentials;

4. Rapid modeling and synchronous dense point mapping to establish a true anatomical model;

5. Integrated with a complete electrophysiological recording system, automatic mapping module, ablation control analysis module, and multi-parameter adjustable stimulation module, achieving multifunctional integration;

6. Supports seamless integration with Jinjiang Electronic's self-developed radiofrequency ablation system and pulsed electric field ablation system, providing three-dimensional guided radiofrequency ablation.(RFA)+Pulsed Field Ablation Under 3D Guidance(PFA)A Variety of Comprehensive Solutions.


Company Introduction

Image

Jinjiang ElectronicFounded in 2002, headquartered in Chengdu, Sichuan,It is one of the few companies in China with a complete line of cardiac electrophysiology products, capable of providing coverage for cardiac electrophysiology procedures.Mapping, Modeling, Ablation, ValidationThe full-process equipment and consumables product portfolio.

Since its establishment, the company has gradually built up through independent research and development.3DPFATreatment PlanA product series at the core.

The core products of the company include a magnetic and electric positioning three-dimensional mapping system, cardiac pulse field ablation equipment, and accompanying catheters. In addition, the company's product range covers multi-channel physiological recorders, radiofrequency ablation devices, radiofrequency ablation catheters, and access consumables, among other cardiac electrophysiology products. The product portfolio is relatively comprehensive, providing complete solutions for patients with tachyarrhythmia. Meanwhile, the company’s product line extends to the field of pain management, meeting more diversified clinical needs.

Financing History

  • 2021Year12To2022Year8During the month, Jinjiang Electronic completedAWheel andBRound of financing, with cumulative financing exceeding7RMB billion, with investors including Hillhouse Capital, Xinda Kunpeng, Guanghua Wutong, Yuanheng Lizhen, Weigao Group, Huimei Capital, etc. This isThe Largest Scale Primary Market Transaction in the History of China's Electrophysiology Industry

  • 2022Year6Month12On [date], Jinjiang Electronic submitted its prospectus to the STAR Market, planning to raise funds.26.91100 million yuan